Indicated if one or several bands are excluded. This will violate the A-weighted result of the measurement unless the band is more than 10 dB below the highest A-weighted band. Please consult the standard.
Indicated if one or several bands are excluded. This will violate the A-weighted result of the measurement unless the band is more than 10 dB below the highest A-weighted band. Please consult the standard.
This warning is issued if a band has a measured F2 > Ld.
This may be avoided by proper calibration, and avoiding measuring a highly reactive sound field. Try to move away from the source or eliminate extranous noise sources to avoid the warning.
This warning is issued if a band has a measured F2 > Ld.
This may be avoided by proper calibration, and avoiding measuring a highly reactive sound field. Try to move away from the source or eliminate extranous noise sources to avoid the warning.
The yellow warning is issued for a segment. The measurement is still compliant to the standard if the summed bands in total scope fullfills the requirement. Otherwise, a red warning will be issued.
If the measurement duration is not satisfying BT ≥ 400, this warning will be issued.
B is the bandwidth of the lowest frequency band in calculated range.
T is the measurement duration.
The temporal variability indicator F1 is > 0.6 for a band. Take actions according to Action E
The extranous intensity criterion is that F3 - F2 < 3 dB for one or more bands.
Try to shield the device under test for directive extraneous noise sources.
One or more bands have negative direction intensity/sound power.
Try to shield the device under test from extraneous noise sources.
Too few measurement positions for the sound field. Criterion 2, N > CF42, is not satisfied. Please see the suggested action code.
The Weighted Sum Frequency range is not set to 50 Hz to 6.3 kHz.
The filter bandwidth is not set to minimum the range 31.5 Hz to 10 kHz.
Indicates that a frequency band outside the Weighted Sum Frequency Range is less than 6 dB below the total A-weighted value when A-weighted.
The standard states that the uncertainty may not be guaranteed in this case.
Indicates that the A-weighted power from bands [50 Hz to 315 Hz] + 6.3 kHz is larger than the A-weighted power for bands 400 Hz to 5 kHz.
The uncertainty for total A is then invalid, and individual band uncertainties apply.
Indicates that the sound intensity probe has not been phase corrected within the last 24 hours.