To perform a measurement according to the standard, set up the parameters in the Measurment menu. The relevant parameters are:
- Averaging time:
The number of seconds per period of measurement.
- Number of averages:
The number of periods to evaluate.
ISO 9614-1 recommends an averaging time of 8 to 12 seconds and 10 periods. Consult the standard for further information about stationarity.
To make a measurement, find a typical and representative position on the hypothetical surface and position the probe there. On the instrument, navigate into the corresponding surface and press and hold on the selected segment to bring up the context menu. Then select Temporal Variability. A special mode is now activated with the current settings. Press the Start button to make the measurment. The instrument will stop automatically when the predefined number of periods have been measured.
Evaluate the measurement by pressing WIEW, the F1 is displayed. Warnings will be given if the measurement is not within the requirement of the standard.
Press accept to store the measurement in the project and populate the F1 indicator for the project.