Before any calibration can be done, the probe must be dismantled. For the Nor1290 probe, please follow the link below for instructions:
Before any calibration can be done, the probe must be dismantled. For the Nor1290 probe, please follow the link below for instructions:
First, make sure the intensity probe is selected as the input channel. Press to enter the calibration menu. Set up the microphone spacer (12 mm recommended). More information:
Input values of the measurement environment:
- Ambient temperature
- Ambient pressure
Back in the calibration menu, tap to enable the phase correction mode. Insert both microphones into the Nor1294 Phase Calibration Coupler. Be careful to make sure the ventilation is open when inserting the microphones, and ensure both microphones are completely inserted before closing the ventilation.
Connect the Nor1294 to the headset connector on the Nor150 (picture below).